In NHS and adult social care settings, a cyber incident on a health IT system is also a patient safety event, yet clinical safety files and security risk registers are usually written by different people who never compare notes. DCB0129 and DCB0160 are the NHS clinical risk management standards for health IT: DCB0129 for manufacturers, DCB0160 for the organisations that deploy and use the systems. This guide explains what the standards require, how their legal footing has strengthened since 2025, where the 2026 review stands, and how to bring cyber threats into the hazard log and clinical safety case so that safety and security assessments reinforce each other. It is for Clinical Safety Officers (CSOs), digital and security leads in trusts, ICBs and care providers, and supplier teams preparing evidence. For the procurement baseline that sits alongside these standards, see the NHS DTAC Guide.
Status line: checked 27 November 2026. The current published releases are DCB0129 (Amd 24/2018) and DCB0160 (Amd 25/2018, effective from 1 July 2018 and listed as Active on the NHS standards directory). NHS England's public consultation on revising them ran from 29 June to 11 September 2026; no revised standard had been published when this guide was checked.
Key takeaways
- Two standards, two audiences. DCB0129 applies to manufacturers of health IT systems; DCB0160 applies to health and care organisations that deploy, use, maintain or decommission them.
- The legal duty has hardened. Section 95 of the Health and Care Act 2022 (in force 7 July 2025) moved information standards from "have regard to" to "must comply", and section 121 of the Data (Use and Access) Act 2025 (in force 5 February 2026) allows IT providers to be brought under a compliance duty.
- The CSO is the accountable signatory. NHS England says the CSO must be a senior clinician with current registration with a professional body, such as the GMC or NMC, and sufficient clinical safety training.
- Cyber belongs in the hazard log. Loss of availability, corrupted data and unauthorised change are clinical hazards with clinical consequences; the Synnovis attack in June 2024 led to more than 10,000 postponed outpatient appointments in south east London.
- Revision is coming. NHS England's review cites AI and growing cyber threats as drivers and reports support for a risk-based, tiered approach, so build processes that will carry over rather than rewriting documents now.
What the two standards cover
| Aspect | DCB0129 | DCB0160 |
|---|---|---|
| Full title | Clinical Risk Management: its Application in the Manufacture of Health IT Systems | Clinical Risk Management: its Application in the Deployment and Use of Health IT Systems |
| Who must apply it | Organisations that develop and maintain health IT systems | Health and care organisations responsible for deployment, use, maintenance or decommissioning |
| Current release | Amd 24/2018 | Amd 25/2018 (published 7 June 2018, effective 1 July 2018) |
| Core outputs named in NHS England guidance | Hazard log, clinical safety case report shared with the deploying organisation, clinical risk management plan | Local risk assessment of how the product will be configured and used, hazard log, clinical safety case, records of the process |
| Named role | Clinical Safety Officer, who reviews and signs off the clinical safety case | Clinical Safety Officer for the deploying organisation |
| Medical devices | Scope extended in 2018 to include medical devices that use software, following the EU Medical Devices Regulation (2017/745) | |
NHS England's adopter guidance says deploying organisations should confirm the developer complies with DCB0129, complete a clinical risk assessment, and not adopt technology that does not meet the standards. Its GP-focused guidance adds that the standards apply beyond catalogue purchases to subcontracted services, hosting and software providers, and record scanning or archiving services where patient data is handled, and that a practice stays responsible for safe operation even when an ICB or national programme has assessed the product.
Legal footing and the 2026 review
- Section 250 of the Health and Social Care Act 2012: both standards are published information standards under this section, announced through Information Standards Notices.
- Stronger compliance powers: NHS England's review supporting information says section 95 of the Health and Care Act 2022 changed the duty to "must comply" from 7 July 2025, and section 121 of the Data (Use and Access) Act 2025 extended scope to IT providers from 5 February 2026. It also says that, for now, bodies with a health and care function must still have regard to the current standards, and that future revisions are likely to use the stronger powers. Treat compliance as mandatory in practice.
- The review: NHS England ran 11 focus groups between January and July 2025, then a public consultation from 29 June to 11 September 2026. Areas under consideration include AI and algorithmic systems, complex system interactions, agile development, post-implementation monitoring, a risk-based tiered approach similar to medical device classification, stronger compliance tracking and better CSO guidance. Mandatory clinical risk management and the patient-centred focus are expected to stay.
Because the revised text is not yet published, avoid committing to changes the consultation only discussed. Design processes (hazard identification, change control, incident learning) that will fit whatever structure the revision adopts.
Why cyber threats are clinical hazards
A clinical hazard is a potential source of harm to a patient. Many cyber scenarios produce exactly that, through four routes:
| Cyber scenario | Clinical hazard it creates | Example clinical consequence | Typical controls (safety and security) |
|---|---|---|---|
| Ransomware or outage of a supplier system | Loss of availability of results, records or ordering | Delayed diagnosis, postponed procedures, transfusion delays | Downtime procedures, tested restore, supplier continuity terms |
| Unauthorised modification of data or configuration | Wrong information presented as correct | Wrong dose, wrong patient, missed allergy alert | Integrity checks, change control, privileged access management, audit logging |
| Compromised interface or integration engine | Messages lost, delayed or misrouted between systems | Results not reaching the requesting clinician | Interface monitoring, message reconciliation, service account control |
| Account compromise or shared credentials | Untraceable or inappropriate clinical actions | Orders placed under another clinician's identity | Individual accounts, smartcard or strong authentication, session controls |
The scale is real. NHS England London reported that the ransomware attack on pathology provider Synnovis on 3 June 2024 led to 10,152 postponed acute outpatient appointments and 1,710 postponed elective procedures at King's College Hospital and Guy's and St Thomas' by 26 September 2024 (unvalidated management information). See the Clinical Downtime and EHR Downtime Procedures Guide and the HL7 v2 and Clinical Interface Engine Security Guide.
Bringing cyber into the hazard log and safety case
NHS England's digital clinical safety guidance describes hazard analysis using techniques such as Functional Failure Analysis (FFA), HAZID, the Structured What-If Technique (SWIFT) and fishbone diagrams, followed by systematic risk evaluation and reassessment across the product's life. Cyber threats fit into that method without a separate process.
- Invite security to hazard workshops: include the security lead or a supplier security architect in HAZID or SWIFT sessions, and add prompts such as "what if this system is unavailable for 14 days?", "what if data is silently altered?" and "what if an interface account is misused?".
- Record cyber causes against clinical hazards: in the hazard log, add cyber causes (ransomware, malicious change, credential misuse) to existing hazards such as "results unavailable" rather than creating a separate cyber register nobody reads.
- Cross-reference controls both ways: where a hazard relies on a security control (backups, MFA, interface monitoring), cite the control owner and evidence; where a security risk has clinical impact, cite the hazard ID in the security risk register.
- Test the controls the safety case relies on: downtime drills, restore tests and failover checks are clinical safety evidence as much as security evidence. Record results in the safety case.
- Re-assess on change: a version upgrade, a new integration, a hosting move or a major security patch can change hazards. Route changes through clinical safety review as part of change management. See the Change Management Security guide.
- Close the loop on incidents: NHS England guidance says a safety incident log must be kept and significant incidents reported to the manufacturer, the commissioner and nationally through the Learn from Patient Safety Events (LFPSE) service. Log cyber-caused clinical incidents there as well as in security incident records.
Manufacturer and deployer responsibilities
| Activity | Manufacturer (DCB0129) | Deploying organisation (DCB0160) |
|---|---|---|
| Hazard identification | Product hazards across design and development | Local hazards from configuration, workflows, integrations and users |
| Cyber input | Secure development, vulnerability handling, security of hosted service | Network placement, identity and access, backup and downtime, supplier management |
| Evidence shared | Clinical safety case report and hazard log extracts | Local safety case referencing the manufacturer's report |
| Change | Release notes with safety impact | Local re-assessment before deploying the change |
| Incidents | Investigate product causes, inform customers | Report to manufacturer, commissioner and LFPSE |
How the standards sit with DTAC, DSPT and device regulation
- DTAC: the Digital Technology Assessment Criteria cover clinical safety, data protection, technical security and interoperability at procurement. The live DTAC guide notes DTAC does not replace DCB0129 and DCB0160. See the NHS DTAC Guide.
- DSPT: the Data Security and Protection Toolkit is the organisational security assurance; use its evidence (backups, incident response, supplier assurance) as inputs to the safety case. See the NHS Data Security and Protection Toolkit guide.
- ISO 14971 and medical devices: NHS England's AI and Digital Regulations Service says completing ISO 14971 does not replace DCB0129 where DCB0129 applies, and that ISO 14971 addresses broader device safety including cyber security. See the Medical Device Cybersecurity Rules Compared guide.
- ISO/IEC 80001-1:2021: NHS England adopter guidance recommends it when integrating a medical device into IT infrastructure, covering safety, effectiveness and security in implementation and use of connected devices.
Where identity fits
Hazards such as "order placed under the wrong clinician" and "interface account used to inject messages" are identity hazards. Smartcard and role-based access, individual rather than shared accounts, and tight control of interface and service accounts are controls a safety case can cite. See the NHS Spine, smartcards and NHS identity security guide and the Machine Identity in Healthcare guide.
Choosing an approach by organisation size
- Acute trusts and large providers: run joint clinical safety and security review boards for significant changes, keep one hazard log per major system with cyber causes, and schedule restore and downtime tests as safety evidence.
- GP practices and small care providers: rely on the manufacturer's safety case and ICB support, but complete a short local assessment covering local configuration, downtime arrangements and who holds administrator rights.
- Manufacturers and SaaS suppliers: include security architects in hazard analysis and state in the clinical safety case report which security controls the customer must operate for the safety claims to hold.
Detection and measurement
- Coverage: percentage of clinical systems with a current DCB0160 safety case that includes cyber causes.
- Change review rate: share of significant changes (upgrades, integrations, hosting moves) that received clinical safety review before go-live.
- Downtime readiness: date of the last tested downtime procedure and restore per critical system, against a target interval.
- Incident linkage: number of security incidents with clinical impact that were also logged in the safety incident log and LFPSE.
- Open cyber-caused hazards: count of hazard log entries with a cyber cause whose residual risk is above the organisation's acceptance level, with an owner and a review date for each.
- Supplier evidence currency: percentage of suppliers with an in-date DCB0129 clinical safety case report on file.
Framework mapping
| Framework | Reference | Relevance |
|---|---|---|
| NIST CSF 2.0 | ID.RA-04, ID.RA-05 | Recording impacts and likelihoods, including clinical impact, to prioritise risk response |
| NIST CSF 2.0 | GV.SC-07 | Supplier risk across the relationship, matching DCB0129 evidence reviews |
| NIST CSF 2.0 | RC.RP-01 | Executing recovery, tied to downtime and restore hazards |
| NIST SP 800-53 Rev 5 | RA-3, CM-4, CP-2, SA-9, SR-6 | Risk assessment, impact analysis of changes, contingency planning, external services and supplier assessments |
| ISO/IEC 27001:2022 Annex A | A.5.19 to A.5.22, A.5.29, A.5.30, A.8.32 | Supplier security, security during disruption, ICT readiness for business continuity, change management |
Common failure patterns
- Separate worlds: a clinical safety case that never mentions cyber, and a security risk register that never mentions patients.
- Safety case written once: no re-assessment after upgrades, new integrations or hosting changes.
- Untested assumptions: the safety case claims "paper downtime procedures mitigate loss of system" with no drill on record.
- Accepting supplier evidence unread: filing a DCB0129 report without checking which customer-side controls it assumes.
- Treating DTAC as sufficient: a passed DTAC assessment does not replace a local DCB0160 assessment.
- Security patches treated as outside clinical safety: urgent patches skip review entirely, or the opposite, clinical sign-off delays a critical security fix; agree a fast-track route in advance so both risks are weighed in hours, not weeks.
- Cyber incidents missing from safety learning: outages handled as IT incidents only, so clinical harm is never reviewed or reported through LFPSE.
Practitioner checklist
- Confirm a named, registered clinician is appointed as CSO with clinical safety training, for each manufacturer and deploying organisation.
- Obtain and read the DCB0129 clinical safety case report for every clinical system, noting the customer controls it assumes.
- Add cyber causes (unavailability, integrity loss, credential misuse, interface compromise) to hazard logs for critical systems.
- Cross-reference hazard IDs in the security risk register and control owners in the hazard log.
- Route upgrades, integrations, hosting moves and major patches through clinical safety review.
- Test downtime procedures and restores and record the results as safety case evidence.
- Log cyber-caused clinical incidents in the safety incident log and report significant ones through LFPSE.
- Track the NHS England review and plan to map existing files to the revised standards when published.
Standards and references
- NHS England: DCB0160 Clinical Risk Management, Deployment and Use of Health IT Systems
- NHS England: National review of clinical risk management standards DCB0129 and DCB0160, supporting information
- NHS England: National review consultation
- NHS England: What is digital clinical safety assurance?
- NHS AI and Digital Regulations Service: Complying with NHS digital clinical risk management standards
- NHS AI and Digital Regulations Service: DCB standards vs ISO 14971
- NHS England London: Update on cyber incident clinical impact in south east London
Related NHI Mgmt Group resources: NHS DTAC Guide · NHS Data Security and Protection Toolkit · Healthcare Incident Response Guide · Healthcare Cyber Regulation Map. This guide summarises the DCB0129 and DCB0160 standards and is not legal advice.