21 CFR Part 11 is the US Food and Drug Administration rule that decides when electronic records and electronic signatures can stand in for paper and ink, and most of its requirements are identity and access controls. Unique signatures, verified identities, authority checks, time-stamped audit trails, password and token management and signature to record linking are all written into the rule. For security and identity teams in pharmaceutical, biotech, medical device, clinical research and food businesses, Part 11 is where the general practice of identity governance meets a regulator who inspects it.
This guide walks through the rule section by section, explains how FDA's 2003 scope and application guidance narrows enforcement, and maps the controls to modern identity, logging and automation, including service accounts and AI agents that touch regulated records. For the cryptography behind signatures, see the Digital Signatures Guide.
This page is general information, not legal advice or regulatory advice. Status as of 17 Nov 2026, based on the eCFR text of 21 CFR Part 11 (up to date as of 23 October 2026) and FDA guidance pages. Confirm with your quality and regulatory affairs functions.
Key takeaways
- Part 11 applies to records that FDA predicate rules (such as GMP, GLP, GCP and quality system rules) require, when those records are kept or submitted electronically (11.1(b)).
- Closed systems need eleven groups of controls in 11.10, including validation, limited access, secure computer-generated time-stamped audit trails, authority checks and device checks.
- Each electronic signature must be unique to one individual and never reused or reassigned (11.100(a)); identity must be verified before a signature is issued (11.100(b)).
- Organisations must certify to FDA, in a letter with a handwritten signature, that their electronic signatures are the legally binding equivalent of handwritten signatures (11.100(c)).
- Non-biometric signatures need at least two distinct components, such as an identification code and a password (11.200(a)).
- FDA's 2003 guidance exercises enforcement discretion on validation, audit trails, copies, retention and legacy systems, but access limits, authority checks, device checks, accountability policies and all signature controls remain enforced.
Part 11 at a glance
| Section | Topic | Security relevance |
|---|---|---|
| 11.1, 11.2 | Scope and implementation | Which records and submissions are in scope |
| 11.3 | Definitions | Closed and open systems, electronic and digital signatures, biometrics |
| 11.10 | Controls for closed systems | Validation, access, audit trails, authority and device checks, training, accountability, documentation control |
| 11.30 | Controls for open systems | Adds encryption and digital signature standards |
| 11.50 | Signature manifestations | Printed name, date and time, meaning of the signature |
| 11.70 | Signature and record linking | Signatures cannot be cut, copied or transferred |
| 11.100 | General signature requirements | Uniqueness, identity verification, certification to FDA |
| 11.200 | Signature components | Two components, session rules, collaboration needed to misuse |
| 11.300 | Identification codes and passwords | Uniqueness, periodic review, loss management, misuse reporting, token testing |
Scope: which records and systems
Section 11.1(a) sets the purpose: criteria under which FDA considers electronic records, electronic signatures and handwritten signatures executed to electronic records to be trustworthy, reliable and generally equivalent to paper records and handwritten signatures on paper. Section 11.1(b) applies the part to records in electronic form that are created, modified, maintained, archived, retrieved or transmitted under any records requirement in FDA regulations, and to records submitted to FDA under the Federal Food, Drug, and Cosmetic Act and the Public Health Service Act. Paper records that are merely transmitted electronically are not covered.
The underlying record requirements are called predicate rules. FDA's guidance cites examples such as 21 CFR 211.180 (drug manufacturing record retention), 820.70(i) (software validation in device quality systems) and 58.130(e) (laboratory study data). Paragraphs (f) onwards of 11.1 carve out records required under certain other parts, including some food safety rules, so the first job is always to list which predicate rules apply to which systems.
Section 11.1(e) requires computer systems, controls and related documentation to be readily available for, and subject to, FDA inspection. Section 11.2 lets organisations replace paper with electronic records and signatures for records they keep, and for submissions where the submission type is listed as acceptable in FDA's public docket.
The definitions in 11.3 drive the control set. A closed system is one where access is controlled by the persons responsible for the content of the records. An open system is one where it is not. An electronic signature is a computer data compilation of symbols executed, adopted or authorised by an individual to be the legally binding equivalent of a handwritten signature. A digital signature is an electronic signature based on cryptographic methods that allows the signer's identity and the data's integrity to be verified.
Closed system controls in 11.10
Persons using closed systems must employ procedures and controls designed to ensure the authenticity, integrity and, when appropriate, confidentiality of electronic records, and to ensure the signer cannot readily repudiate the signed record. The controls are:
- (a) Validation to ensure accuracy, reliability, consistent intended performance and the ability to discern invalid or altered records.
- (b) Accurate and complete copies in human-readable and electronic form for FDA inspection, review and copying.
- (c) Protection of records so they can be accurately and readily retrieved throughout the retention period.
- (d) Limiting system access to authorised individuals.
- (e) Secure, computer-generated, time-stamped audit trails that independently record the date and time of operator entries and actions that create, modify or delete records. Changes must not obscure previously recorded information, and audit trails must be retained at least as long as the records and be available for FDA review.
- (f) Operational system checks to enforce permitted sequencing of steps and events.
- (g) Authority checks so only authorised individuals can use the system, sign a record, access the operation or computer system input or output device, alter a record, or perform the operation at hand.
- (h) Device checks to determine the validity of the source of data input or operational instruction.
- (i) Education, training and experience for people who develop, maintain or use the systems.
- (j) Written policies that hold individuals accountable and responsible for actions initiated under their electronic signatures, to deter record and signature falsification.
- (k) Controls over systems documentation, including distribution, access and use of operation and maintenance documentation, and revision and change control with a time-sequenced audit trail.
For open systems, 11.30 requires the same controls as appropriate, plus additional measures such as document encryption and appropriate digital signature standards, to ensure authenticity, integrity and confidentiality from creation to receipt. Any regulated record that travels over the public internet to a partner, contract organisation or cloud service where you do not control access should be assessed against 11.30.
Electronic signature requirements
Manifestation and linking (11.50 and 11.70)
Signed electronic records must show the printed name of the signer, the date and time of signing, and the meaning of the signature, such as review, approval, responsibility or authorship (11.50(a)). These items are subject to the same controls as the record and must appear in any human-readable form such as a display or printout (11.50(b)). Section 11.70 requires signatures to be linked to their records so they cannot be excised, copied or otherwise transferred to falsify a record by ordinary means. Cryptographic binding is the strongest way to meet this; database foreign keys alone are weaker if administrators can edit them.
Uniqueness, identity and certification (11.100)
Each electronic signature must be unique to one individual and must not be reused by, or reassigned to, anyone else (11.100(a)). Before an organisation establishes, assigns, certifies or otherwise sanctions a signature or any element of one, it must verify the identity of the individual (11.100(b)). Persons using electronic signatures must, before or at the time of use, certify to FDA that the electronic signatures in their system, used on or after 20 August 1997, are intended to be the legally binding equivalent of traditional handwritten signatures (11.100(c)). The certification is submitted with a handwritten signature, and FDA may ask for further certification or testimony about a specific signature.
Components and controls (11.200)
Signatures not based on biometrics must use at least two distinct identification components such as an identification code and password (11.200(a)(1)). Within a single continuous period of controlled access, the first signing uses all components and later signings use at least one component that is only executable by, and designed to be used only by, the individual. Signings outside a continuous session require all components each time. Signatures must be used only by their genuine owners, and must be administered so that attempted use by anyone else requires the collaboration of two or more individuals (11.200(a)(2) and (3)). Biometric signatures must be designed so they cannot be used by anyone other than their genuine owners (11.200(b)).
Codes, passwords and tokens (11.300)
- (a) No two individuals may have the same combination of identification code and password.
- (b) Code and password issuances must be periodically checked, recalled or revised.
- (c) Loss management procedures must deauthorise lost, stolen, missing or compromised tokens, cards and devices, and issue replacements under suitable controls.
- (d) Transaction safeguards must prevent unauthorised use, and attempts must be reported immediately and urgently to the security unit and, as appropriate, management.
- (e) Tokens and cards must be tested initially and periodically to ensure they function properly and have not been altered.
Note how 11.300(b) predates current password guidance. Periodic review satisfies it; it does not force arbitrary password expiry, and phishing resistant authenticators can be documented as meeting the component and loss management rules.
Enforcement discretion and FDA guidance
FDA's September 2003 guidance, Part 11, Electronic Records; Electronic Signatures: Scope and Application, is final and remains FDA's current thinking. It interprets the scope narrowly and states that FDA intends to exercise enforcement discretion for Part 11 requirements on validation, audit trails, record retention and record copying, and for legacy systems operational before 20 August 1997 that still meet predicate rules. Decisions on validation, audit trails and retention should be based on a justified and documented risk assessment.
The same guidance is explicit that Part 11 remains in effect and that FDA will still enforce limiting system access to authorised individuals, operational system checks, authority checks, device checks, training, accountability policies, systems documentation controls, open system controls and all signature requirements in 11.50, 11.70, 11.100, 11.200 and 11.300. Predicate rules are fully enforced, and many predicate rules impose their own audit and data integrity requirements, so discretion on 11.10(e) rarely means no audit trail in practice.
In October 2024 FDA issued final guidance, Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers, for sponsors, investigators, institutional review boards and contract research organisations. Clinical teams should read it alongside the 2003 guidance.
Enforcement happens through inspections and the remedies available under the predicate statutes. Part 11 itself contains no penalty schedule; findings typically appear as inspectional observations and warning letters tied to data integrity and predicate rule failures.
What it means for security and identity teams
Part 11 was written about people, but modern regulated systems are run by integrations, service accounts, robotic process automation and, increasingly, AI agents. Three rules follow.
- Machines do not sign. An electronic signature belongs to one individual (11.100(a)). A service account or agent can create or modify records, but it must never execute a signature. If an agent prepares a batch record for approval, a named human signs it, with the 11.50 meaning recorded.
- Machines are still users for 11.10. Limited access (d), authority checks (g) and device checks (h) apply to every identity that can change regulated records. Each integration needs its own identity, least privilege scoped to the operations it performs, and an owner. Shared or generic accounts break attribution and the audit trail. See the Service Account Security Guide.
- The audit trail must name the actor. An entry that says "system" or "api" for an automated change does not let an inspector reconstruct who or what acted. Record the non-human identity, the human who authorised the job, and the time. The AI Agent Audit Trails for Regulators Guide covers schema design.
Administrator access is the other big theme. Database administrators and platform engineers who can edit records or audit tables outside the application undermine 11.10(e) and 11.70. Use privileged access management, separate duties between system owners and administrators, and route direct data fixes through change control with their own audit trail.
Compliance steps
- Inventory systems against predicate rules. For each system, record which predicate rules apply, whether it is closed or open, and whether it uses electronic signatures.
- Risk assess. Document a risk assessment per system covering product quality, safety and record integrity, as the 2003 guidance expects, and decide validation depth and audit trail scope from it.
- Govern identities. Verify identity before issuing signatures, enforce unique accounts, run periodic access reviews, and give every service account and agent an owner and scoped rights.
- Harden audit trails. Make them computer generated, time-stamped from a trusted time source, append only, retained as long as the records, and readable by inspectors. Protect them from administrator edits.
- Implement signature controls. Two components, session rules, manifestation and cryptographic or equivalent linking; submit the 11.100(c) certification letter.
- Manage tokens and passwords. Loss procedures, misuse reporting to the security unit, and periodic testing of tokens and cards.
- Prepare for inspection. Keep validation evidence, procedures, training records and accountability policies readily available under 11.1(e).
Common failure patterns
- Shared laboratory or shop floor logins, which break 11.100(a) and 11.300(a) and make audit trails meaningless.
- Audit trails that administrators can disable or edit, or that are switched off for performance.
- Service accounts with broad write access to regulated tables and no named owner.
- Automated jobs or AI agents applying approvals under a human's credentials.
- Signature records missing the meaning of the signature, or printouts that drop the 11.50 elements.
- Clock drift between systems, producing time stamps that cannot be reconciled during an investigation.
- No 11.100(c) certification letter on file with FDA.
Practitioner checklist
- Every system holding Part 11 records is listed with its predicate rules and closed or open status.
- Each human and non-human identity is unique, owned and reviewed at least periodically.
- No service account or AI agent can execute an electronic signature.
- Audit trails capture create, modify and delete with actor, date and time, and preserve prior values.
- Audit trails are protected from privileged users and retained as long as the records.
- Signatures display name, date and time and meaning, and are bound to their records.
- Lost or compromised authenticators are revoked under a written procedure, and misuse attempts reach the security unit.
- The 11.100(c) certification has been submitted and a copy is kept.
Standards and references
- eCFR: 21 CFR Part 11, Electronic Records; Electronic Signatures
- FDA: Part 11, Electronic Records; Electronic Signatures: Scope and Application (2003 guidance)
- FDA: Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations, Questions and Answers (2024)
- NIST: SP 800-53 Rev. 5 (AU-9 Protection of Audit Information, AU-10 Non-repudiation, IA-2 Identification and Authentication)
- NIST: SP 800-63B-4 Authentication and Authenticator Management
- NIST: SP 800-63A-4 Identity Proofing and Enrollment
- ISO: ISO/IEC 27001:2022 Information security management systems
Related NHI Mgmt Group resources: Digital Signatures Guide · AI Agent Audit Trails for Regulators Guide · AI Agent Attribution and Non-Repudiation · HIPAA Security Rule Guide