Medical device makers face FDA and EU reviewers who expect a documented secure development lifecycle, and the Health Sector Coordinating Council's Joint Security Plan (JSP2) is the sector's own blueprint for building one. JSP2 is version 2.0 of the Medical Device and Health IT Joint Security Plan, published in March 2024 by the HSCC Cybersecurity Working Group's Medtech Task Group. It is voluntary: the document says it is not a regulatory document and not a standard. This guide explains how a product security team turns its six lifecycle phases into working processes, how those phases line up with the FDA premarket guidance and MDCG 2019-16, and where a small manufacturer should start. For the wider regulatory picture, see the Medical Device Security Guide.

Key takeaways

  • JSP2 is a framework, not a rule. It carries no legal force on its own, but FDA's February 2026 premarket guidance names it as an example of a secure product development framework (SPDF), so following it is a credible way to show reviewers a structured process.
  • Six phases cover the whole product life. Concept, Risk Management, Supplier Management, Design and Development, Verification and Validation, and Maintenance, each with defined activities and outputs.
  • It plugs into your quality system. JSP2 asks manufacturers to map security activities into existing ISO 13485 and design control processes rather than run a parallel programme.
  • Postmarket expectations are concrete. For medium to critical vulnerabilities JSP2 recommends notifying customers within 30 days and making a fix available within 60 days, with standard and out-of-cycle update tracks.
  • Customer documentation is part of the product. Appendix D sets out what to ship: an MDS2 form, an SBOM, end-of-life and end-of-support dates, a patch management plan and a residual risk summary.

What JSP2 is and who wrote it

The first Joint Security Plan appeared in January 2019. Version 2.0 was drafted by a committee formed in 2022, co-chaired by representatives from a large device manufacturer, a medtech security consultancy and the US Food and Drug Administration, with contributors from industry, FDA and consulting firms. The HSCC is the private sector half of the US healthcare and public health critical infrastructure partnership, so JSP2 reflects consensus among manufacturers, regulators and hospitals rather than the view of one body.

The plan is written to be applied across a product portfolio, globally, and at any company size, with rigour scaled to the complexity and risk of each product. It draws heavily on recognised standards: IEC 81001-5-1 for lifecycle security activities, ANSI/AAMI SW96:2023 for security risk management, ISO 14971:2019 and ISO/TR 24971:2020 for safety risk, ISO 13485:2016 for quality management, and ISO/IEC 29147 and 30111 for vulnerability disclosure and handling. Where those standards say what must be achieved, JSP2 adds practical example outputs and appendices, such as example design input requirements, example customer documentation and an example organisational structure.

  • Who should use it: manufacturers of networked or software-driven devices and health IT vendors, especially product security, quality and regulatory affairs teams.
  • Who it is not for: hospitals securing devices they already own. Healthcare delivery organisations should look at the MDS2 and Medical Device SBOM Guide for Hospitals and the HSCC's Health Industry Cybersecurity Practices instead.
  • What changed from v1: v2 adds a dedicated supplier management phase, deeper SBOM and exploitability guidance, and replaces the old CMMI-style maturity questions with participation in the Medical Device Innovation Consortium (MDIC) benchmarking survey.

Lifecycle phases

JSP2 organises product security around the total product lifecycle (TPLC). Each phase below lists the key activities and the evidence a reviewer or auditor would expect to see.

PhaseKey activitiesTypical outputs
A. ConceptRelease and change planning with security resourcing; voice-of-customer security needs; security management planning including risk acceptability criteriaSecurity Management Plan; security needs captured as design inputs
B. Risk ManagementSecurity design risk assessment; feeding security risks with safety impact into the ISO 14971 safety risk file; treating design risks and known vulnerabilities as separate categoriesSecurity Risk Management Summary Report approved by management
C. Supplier ManagementSecurity assessment before purchase; contract terms; licence review; ongoing vulnerability monitoring; SBOM review; incident notification; periodic supplier risk reviewSupplier security requirements in contracts; supplier SBOMs per release
D. Design and DevelopmentSecurity requirements; threat modelling of the architecture; secure coding, code review, static and dynamic analysis; patch and update planningThreat model; security requirements traced to design; update plan
E. Verification and ValidationVerifying each security control with traceable, ideally automated tests; scanning for known vulnerabilities against NVD and the CISA KEV catalogue; penetration, robustness (including fuzz) and resiliency testingTest reports; known vulnerability assessment re-rated in product context
F. MaintenanceSurveillance of disclosure reports, ISAOs and supplier notices; vulnerability and end-of-life management; incident response with hospitals; patch deployment; customer communicationCVD policy and PSIRT procedure; security bulletins; end-of-support notices

Two points are easy to miss. First, JSP2 asks for exploitability to be reassessed in the product's own context, not copied from a generic CVSS score; its appendices cover exploitability assessment methods and reference the MITRE CVSS rubric for medical devices and CISA's SSVC guide. Second, it separates standard update cycles from critical out-of-cycle updates and suggests cycle frequencies by device type, noting that implanted devices are updated rarely and have no regular cycle. See the Implantable and Home-Use Medical Device Security Guide for that class.

Roles and governance

Section VIII of JSP2 asks for an accountable leader with adequate resources and for defined roles, responsibilities and training needs. Larger companies may run a dedicated product security function; smaller ones can assign the role to an existing quality or engineering lead, provided accountability is explicit.

  • Accountable executive: owns risk acceptance. JSP2 expects residual security risk to go through a formal acceptance process signed by senior leadership.
  • Product security lead or team: runs threat modelling, security requirements, testing oversight and the PSIRT.
  • Quality and regulatory affairs: embed security activities in design controls, CAPA, complaint handling and submissions so evidence is produced as part of normal work.
  • Procurement and supplier quality: carry the supplier management phase, including contract language for SBOMs and vulnerability notification timelines.
  • Service and customer support: deliver patches, field questions from hospital security teams and coordinate incident response.

Training is ongoing, with needs reassessed periodically. The HSCC also publishes a Medtech Vulnerability Communications Toolkit and model contract language, which JSP2 references for supplier and customer agreements.

Mapping to FDA and EU MDR expectations

JSP2 does not create obligations, but it produces the evidence both main regimes ask for. FDA's guidance Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions, reissued on 3 February 2026 to replace the June 2025 version, describes an SPDF as processes that reduce the number and severity of vulnerabilities across the device lifecycle and lists JSP2, IEC 81001-5-1 and ANSI/ISA 62443-4-1 as example frameworks. In the EU, cybersecurity duties sit in MDR Annex I sections 17.2, 17.4, 18.8 and 23.4(ab), and MDCG 2019-16 rev. 1 (July 2020) explains how to meet them. Appendix G of JSP2 maps its activities to MDCG 2019-16 by section.

JSP2 phase or outputFDA premarket guidance (Feb 2026) and section 524BMDCG 2019-16 rev. 1
A Concept: Security Management PlanSPDF integrated with the quality management systemSection 3: secure design and manufacture, security risk management
B Risk ManagementCybersecurity risk assessment with pre and post mitigation scoring, kept separate from but linked to safety riskSection 3: security risk assessment and benefit-risk analysis
C Supplier Management and SBOMMachine-readable SBOM with each component's support level and end-of-support date; required for cyber devices under 524B(b)(3)Section 4.2: SBOM listed as additional security documentation; section 5: monitoring third-party components
D Design: threat model, architectureThreat model; architecture views (global system, multi-patient harm, updateability and patchability, security use cases)Section 2.3 and 3: defence in depth, not relying on the operating environment
E Verification and ValidationSecurity testing including penetration testing, with tester independence and scope documented; list of unresolved anomaliesSection 3: verification and validation, including fuzz and penetration testing
Appendix D customer documentationCybersecurity labelling scaled to the intended userSections 4.2 and 4.3: instructions for use and information for healthcare providers, MDS2
F Maintenance: CVD, patching, end of supportCybersecurity management plan; 524B(b)(1) plans to monitor, identify and address postmarket vulnerabilities including coordinated disclosureSection 5: post-market surveillance and vigilance; serious incident reporting

Detail on what each regulator expects is covered in the FDA Medical Device Cybersecurity Guide and the EU MDR Cybersecurity and MDCG 2019-16 Guide; for how the rules differ market by market, see Medical Device Cybersecurity Rules Compared.

SBOM and vulnerability disclosure duties

  • What JSP2 recommends: require SBOMs from suppliers in contracts, receive an updated SBOM with every supplier release, monitor SBOM components for new vulnerabilities, and ship an SBOM to customers at purchase and after major upgrades.
  • What is legally required: in the US, section 524B of the FD&C Act makes an SBOM mandatory in premarket submissions for cyber devices. In the EU, the MDR does not name SBOMs, but MDCG 2019-16 lists one among expected security documentation. Treat JSP2's SBOM practice as the way to meet both.
  • Disclosure: JSP2 expects a published coordinated vulnerability disclosure policy and a PSIRT procedure aligned to ISO/IEC 29147 and 30111, and customer bulletins that include CVSS scores and patient safety information. The 30 day notification and 60 day fix targets are JSP2 recommendations, not statutory deadlines; FDA's postmarket guidance sets its own reporting expectations.

See the SBOM Guide and the AAMI TIR57 and SW96 Medical Device Risk Guide for the underlying practices.

Maturity and metrics

Section X of JSP2 recommends periodic self-assessment against the plan and participation in the MDIC JSP benchmarking survey, whose annual report lets manufacturers compare their progress with the sector. The plan accepts that organisations still maturing may treat some activities as aspirational, provided they work towards them. Manufacturers may share their maturity results with customers and partners.

Where a small manufacturer should start

  1. Name the accountable leader and write a one-page product security policy that references JSP2 and IEC 81001-5-1.
  2. Build the minimum premarket evidence first: threat model, security risk assessment linked to the ISO 14971 file, SBOM and a penetration test report. These are what reviewers ask for.
  3. Publish a CVD policy and set up a monitored intake route, even if the PSIRT is two people.
  4. Add supplier clauses for SBOM delivery and vulnerability notification to every new software or component contract, using the HSCC model contract language.
  5. Write the customer pack (MDS2, SBOM, end-of-support date, patch plan) once and version it with each release.
  6. Self-assess annually against the six phases, record gaps as CAPA or improvement items, and expand automation (scanning, SBOM monitoring) as the portfolio grows.

Larger manufacturers should go further: portfolio-wide SBOM monitoring, a standing PSIRT with defined service levels, and benchmarking every year.

Metrics that show it is working

MetricWhy it matters
Percentage of medium to critical vulnerabilities where customers were notified within 30 daysTracks the JSP2 communication target
Percentage with a fix available within 60 daysTracks the JSP2 remediation target
Percentage of released products with a current SBOM and stated end-of-support dateRequired evidence for cyber devices and for customers
Percentage of software suppliers under contracts with SBOM and notification clausesShows supplier management phase coverage
Security findings per submission raised by reviewersA falling count indicates the SPDF evidence is complete
Phases rated fully implemented in the annual self-assessmentTracks maturity over time

Framework mapping

JSP2 areaNIST CSF 2.0NIST SP 800-53 Rev. 5
Governance and accountable leaderGV.RR-01SA-15
Risk managementID.RA-01RA-3
Supplier managementGV.SC-05SR-3, SR-6
Design and developmentPR.PS-06SA-8
Verification and validationID.RA-01SA-11, RA-5
Maintenance and incident responseRS.MA-01, ID.IM-01SI-2

Common failure patterns

  • Treating JSP2 as a certificate. There is no JSP2 certification. Claiming "JSP2 compliant" in a submission without the underlying evidence helps nobody; cite it as the SPDF you follow and show the outputs.
  • A parallel security process. Security work run outside design controls produces artefacts that do not trace to requirements or risk files, which is exactly what reviewers check. JSP2 explicitly asks for integration with quality processes.
  • Generic CVSS scores. Copying a library's base score without assessing exploitability in the device context leads to either alarm fatigue or missed patient safety risk.
  • SBOMs that stop at release. An SBOM produced for the submission and never monitored gives no postmarket value; JSP2 expects ongoing monitoring and supplier updates.
  • No end-of-support date. Hospitals cannot plan replacement, and FDA expects support status for SBOM components. JSP2 asks for end of life to be communicated before sales stop.
  • Supplier contracts with no security terms. Without notification clauses, a manufacturer may learn of a component vulnerability from a customer.

Practitioner checklist

  • Confirm you are using JSP v2.0 (March 2024) and record it as your SPDF reference in the quality manual.
  • Name an accountable executive and a product security lead, and define residual risk sign-off.
  • Map each of the six phases to existing design control and quality procedures.
  • Produce a threat model, security risk assessment, SBOM with support status and penetration test report for each new submission.
  • Add SBOM delivery and vulnerability notification terms to supplier contracts.
  • Publish a coordinated vulnerability disclosure policy and run a PSIRT procedure aligned to ISO/IEC 29147 and 30111.
  • Define standard and critical update tracks and test them before release.
  • Ship the customer documentation pack: MDS2, SBOM, end-of-support date, patch plan, residual risk summary.
  • Self-assess annually and consider the MDIC benchmarking survey.

Standards and references

Related NHI Mgmt Group resources: Medical Device Security Guide · FDA Medical Device Cybersecurity Guide · EU MDR Cybersecurity and MDCG 2019-16 Guide · MDS2 and Medical Device SBOM Guide for Hospitals · SBOM Guide

PDF for subscribers

Take this guide with you

Every section, table and checklist in one PDF, with links to the sources and related guides. Subscribe to our weekly NHI & AI Identity Journal and we’ll email you a download link.

Free weekly newsletter

Subscribe to the NHI & AI Identity Journal

The latest on NHI and Agentic AI security – articles, research, breaches, news and events every week.

Bonus 33% off our NHI Course when you subscribe.

PDF Downloads for Newsletter Subscribers

Get the guide

Enter your details and we’ll email you a download link.

The full guide is free to read online. The PDF is a thank you for subscribing.

Fill in all fields marked * and tick both boxes to continue.