Most security problems with connected medical devices are bought in: once a device is clinically validated and on the ward, the hospital cannot easily change its operating system, accounts, remote access or support period. The procurement process is the one point where a healthcare delivery organisation has leverage over those things. This checklist gives hospital procurement teams, clinical engineering (biomedical engineering) and IT security a phase-by-phase set of questions, the MDS2 fields that matter most, and model contract terms drawn from the Health Sector Coordinating Council's Model Contract-Language for MedTech Cybersecurity, version 2 (MC2 v2, November 2025). It is a working template, not legal advice: have contract wording reviewed by your own counsel.
How to use this template
- Replace text in square brackets, such as [Organisation], [Clinical engineering lead] and [risk tier], with your own roles and thresholds.
- Apply it to every purchase, lease, loan, donation, trial and renewal of a device that stores, processes or transmits data or connects to any network, including software as a medical device and vendor-hosted cloud components.
- Scale depth to risk (see Implementation guidance). Standalone, non-networked equipment may need only sections 1, 2 and 6.
- Record each answer with its evidence (MDS2 question ID, document reference, contract clause number) in the [procurement record] and in the device inventory entry. See the Hospital Medical Device Inventory and IoMT Visibility Guide.
- Treat a "no", "N/A" without explanation or "unknown" as a finding to accept, mitigate or negotiate, not an automatic rejection. Clinical need may outweigh a security gap, but the decision and the compensating controls must be recorded and owned.
- For the depth behind the MDS2 and SBOM review, see the MDS2 and Medical Device SBOM Guide for Hospitals. This template does not repeat that guide; it turns it into checkpoints.
1. Before tender
Security requirements belong in the specification, not in a questionnaire sent after the clinical team has chosen a product.
- Owner and team: [Organisation] names a procurement lead, a [Clinical engineering lead], an [IT security lead] and, where patient data is involved, the [Data protection officer] before market engagement starts.
- Risk tier: classify the device by connectivity (none, local network, internet or vendor cloud), data handled (none, personal data, health data) and clinical impact if unavailable or manipulated. Record the tier in the procurement record.
- Regulatory status: require evidence of market authorisation for the jurisdiction of use (for example FDA clearance or approval; CE marking under the EU MDR). For devices submitted to FDA since the cyber device requirements took effect, ask whether the device is a "cyber device" under FD&C Act section 524B. See the FDA Medical Device Cybersecurity Guide and the EU MDR Cybersecurity and MDCG 2019-16 Guide.
- Mandatory documents in the tender pack: a current, revision-controlled MDS2 (ANSI/NEMA HN 1-2019), an SBOM in a machine-readable format, network and data flow diagrams, a list of ports and services, and the manufacturer's coordinated vulnerability disclosure policy.
- Support horizon: state the expected service life of the device at [Organisation] (for example [10] years) and require the bidder to state the end of support date for the device and for its operating system and major third-party components.
- Network fit: confirm with [IT security lead] which network zone the device will sit in, and include its requirements (supported authentication, encryption, logging to [SIEM]) in the specification. See the Hospital Clinical Network Segmentation Guide.
- Scoring: give security a weighted score in tender evaluation (for example [15]% for high-tier devices) so that it can change the outcome, and state that a missing MDS2 or SBOM is non-compliant.
2. MDS2 review
The MDS2 form (ANSI/NEMA HN 1-2019) is a standard set of yes/no questions grouped into security capability sections with question IDs such as PAUT-1 or SBOM-1. FDA's premarket guidance notes that an MDS2 and the Joint Security Plan customer security documentation may address many of its labelling recommendations, so it is reasonable to expect one for any connected device. Review it against your requirements, not in isolation.
| MDS2 section | Key questions to check | Answer that should trigger follow-up |
|---|---|---|
| DOC Device description | Coordinated vulnerability disclosure programme (DOC-8); network or data flow diagram available (DOC-10) | No disclosure programme; no diagram |
| AUTH Authorization | Login required (AUTH-1); role-based privilege levels (AUTH-2); can the owner grant itself admin rights (AUTH-3) | Shared logins only; no owner admin control |
| PAUT Person authentication | Unique IDs and passwords for all users and roles (PAUT-1); lockout after failed logons (PAUT-3); multi-factor support (PAUT-8) | Shared or fixed accounts; no lockout |
| CSUP Cybersecurity product upgrades | Software or firmware needing security updates (CSUP-1); customer notified when updates are approved (CSUP-7); automatic installation (CSUP-8) | Updates need a site visit with no stated timeline |
| MLDP Malware detection and protection | Anti-malware supported (MLDP-2); owner can install or update it (MLDP-2.3); host-based intrusion detection or allow-listing (MLDP-5) | No anti-malware and no allow-listing |
| RMOT Remote service | Remote service connections (RMOT-1); remote predictive maintenance data (RMOT-2); other remote functions such as updates (RMOT-3) | Always-on vendor access; method not described |
| SAHD System and application hardening | Hardened to industry standards (SAHD-1); mechanisms to confirm installed software is authorised (SAHD-3.1); owner integrity checks (SAHD-4) | No hardening baseline stated |
| STCF and TXCF Storage and transmission confidentiality | Encryption at rest (STCF-1); data encrypted before transmission (TXCF-2); connections limited to authenticated systems (TXCF-4) | Health data stored or sent unencrypted |
| RDMP Roadmap for third-party components | Secure development process followed (RDMP-1); support dates maintained (RDMP-3); plan for third-party component end of life (RDMP-4) | No support dates or end-of-life plan |
| SBOM Software bill of materials | SBOM available (SBOM-1); standard method to describe components (SBOM-2); update process (SBOM-4) | No SBOM, or SBOM only on request after purchase |
- Version check: the MDS2 must name the exact model and software version being offered and carry a date or revision. A form for a previous software version is not evidence.
- Read the notes: "yes" answers that depend on optional features, extra licences or customer configuration must say so. Record which options you are buying.
- Cross-check: compare the MDS2 against the ports and services list, the SBOM and, at acceptance, the device itself (section 6). Inconsistencies are findings.
- Accounts: every account the MDS2 or manuals reveal, including service and hardcoded accounts, is a non-human identity to inventory and govern. See the Machine Identity in Healthcare.
3. SBOM and patching commitments
Section 524B of the FD&C Act requires sponsors of cyber devices to provide an SBOM that includes commercial, open-source and off-the-shelf components, a plan to monitor and address postmarket vulnerabilities including coordinated vulnerability disclosure, and processes to make updates available on a reasonably justified regular cycle, with critical vulnerabilities that could cause uncontrolled risks addressed as soon as possible out of cycle. FDA's February 2026 premarket guidance adds that the SBOM should be machine-readable and should give, for each component, its level of support and end-of-support date, alongside known vulnerabilities. Ask for that in procurement regardless of jurisdiction.
- SBOM quality: machine-readable (for example SPDX or CycloneDX), covering operating system, commercial, open-source and proprietary components, with version, supplier, level of support and end-of-support date per component.
- SBOM updates: an updated SBOM with every software release, available on a continuing basis (for example through a customer portal).
- Known vulnerabilities: a current list of known vulnerabilities in the SBOM components, with the manufacturer's assessment of exploitability in this device and any compensating controls.
- Patch validation and cadence: who validates third-party patches (operating system, database, browser), how long validation takes, the regular release cycle, and who installs patches (manufacturer, remote, or [Organisation] staff).
- Support period: a written end of support date for the device and its operating system, compared against the service life stated in the tender.
- Vulnerability communications: named contacts, notification channel and service levels for urgent and routine security notices.
4. Contract clauses
MC2 v2 is a model, not a mandate, and its document states that it is not legal advice. Use its clause numbers to anchor negotiation, then adapt timelines to the device's risk tier and your leverage. Group purchasing organisations and framework agreements are often the practical route for smaller hospitals.
| Requirement | MC2 v2 clause | Model position (adapt as needed) |
|---|---|---|
| Accounts and authentication | 6, 16, 29 | Disclose all accounts, authentication methods and any hardcoded passwords or bypasses; remove unneeded accounts; no hardcoded passwords, first-login password change and central directory support where feasible |
| Ports, services and hardening | 4, 7, 11 | Security features on by default or flagged for set-up; unneeded ports, services and software disabled or removed; services documented in the MDS2, manual or SBOM |
| Encryption and logging | 13, 14 | Encrypt sensitive data at rest and in transit; collect, retain and provide timestamped security logs |
| Remote access | 17 | Disclose remote access technology, protocols, ports, encryption, authentication and logging, or follow [Organisation]'s method |
| Vulnerability notification | 23 | Notify uncontrolled risks within 30 days; scoring, remediation and compensating controls within 2 days of notice |
| Patching | 31, 32 | Third-party and supplier patches applied or made available within 30 days of release, on a regular cadence with installation instructions |
| Exploited and KEV-listed vulnerabilities | 33 | Notify within 3 business days of a vulnerability being exploited or added to the CISA Known Exploited Vulnerabilities catalogue |
| Incident notification | 35 | Written notice within 5 days of confirming an incident affecting customer data, unless law directs otherwise |
| Operating system support | 40 | No unsupported OS at delivery; update path at supplier cost within 2 years of OS end of support; notice when an OS is within 1 year of obsolescence |
| SBOM | 44 | SBOM covering open-source, commercial, proprietary and OS components, at least FDA minimum elements; monitor unsupported components |
| Lifecycle notices | 45 | Advance notice of end of life (24 months), end of guaranteed support (24 months) and end of support (36 months), with compensating-control documentation |
| Secure decommissioning | 27 | Wipe or destroy data on returned products in line with NIST SP 800-88, with confirmation on request |
- Jurisdiction: MC2 v2 was written for the US market; clauses such as offshoring (24) and HIPAA record retention need adapting elsewhere. Align data clauses with your data protection law and your data processing agreement.
- Security assessment support: include the right to request scanning or penetration test evidence (clause 42) and complete, accurate questionnaire responses (clause 43).
- Order of precedence: make sure the security schedule is not overridden by the supplier's standard terms or service agreement.
5. Acceptance and onboarding
Do not accept a device into clinical use until it matches what was disclosed and contracted.
- Inventory first: record the device in the asset inventory with model, serial, software version, MDS2 and SBOM references, owner, location, network zone and end of support date.
- Verify the build: compare the installed software version, open ports and running services with the MDS2, SBOM and ports list. Use a passive network capture or an agreed scan in a test segment, coordinated with the manufacturer to avoid disrupting the device.
- Change default credentials: change or disable default and service accounts the contract allows you to manage, store credentials in the [password vault] and record which accounts remain manufacturer-controlled.
- Place it in its zone: connect to the agreed clinical segment with firewall rules limited to documented destinations and ports.
- Set up remote access: route manufacturer access through [Organisation]'s approved remote access method, with per-session approval or time-limited accounts and session logging.
- Turn on logging: forward available security logs to [SIEM] and confirm events arrive.
- Register for notices: subscribe [Clinical engineering lead] and [IT security lead] to the manufacturer's security notification channel and SBOM portal.
- Sign off: clinical engineering, IT security and the clinical owner sign acceptance, listing any accepted findings with compensating controls and a review date.
Implementation guidance
| Risk tier (example) | Typical device | Sections required | Sign-off |
|---|---|---|---|
| Low | No network connection, no stored patient data | 1, 5 (inventory only) | [Clinical engineering lead] |
| Medium | Local network, stores or sends patient data, no vendor remote access | 1 to 5 | Plus [IT security lead] |
| High | Life-sustaining or diagnostic device on the network, vendor remote access or cloud components | All sections, plus acceptance testing in a test segment | Plus [Data protection officer], [Legal] |
- Embed in intake: add a mandatory "does it connect or store data?" question to purchase, lease, loan, trial and donation requests so equipment does not bypass the checklist.
- Pilot on one category: start with a high-volume category such as infusion pumps or patient monitors, refine the template, then extend.
- Keep evidence live: store MDS2, SBOM and contract references against the inventory record, and re-run sections 2 and 3 at every major software release and contract renewal.
- Plan for end of support: track end of support dates as a portfolio and start replacement or compensating-control planning at least [24] months ahead, in line with the notice periods you negotiated.
Measure it: share of connected device purchases that passed through the checklist; share of active devices with a current MDS2 and SBOM on file; share of contracts with vulnerability notification and end-of-support clauses; number of devices past end of support; median days from manufacturer patch release to installation.
| Checklist section | NIST CSF 2.0 | NIST SP 800-53 Rev. 5 |
|---|---|---|
| 1 Before tender | GV.SC-05, GV.SC-06 | SA-4 Acquisition Process; SR-5 Acquisition Strategies, Tools, and Methods |
| 2, 3 MDS2, SBOM and patching | GV.SC-07 | SR-4 Provenance; SA-22 Unsupported System Components |
| 4 Contract clauses | GV.SC-05 | SR-8 Notification Agreements |
| 5 Acceptance and onboarding | ID.AM-08 | SA-4 |
| End of support and disposal | GV.SC-10 | SA-22 |
Common mistakes: accepting an MDS2 for a different software version; asking for an SBOM only after purchase, when leverage has gone; leaving vendor remote access permanently open because it was set up during installation; letting donated, loaned or trial devices skip the process; and agreeing a support period shorter than the device's planned service life without a replacement budget. For wider supplier due diligence, see the Third-Party Risk Management Guide.
Standards and references
- HSCC: Model Contract-Language for MedTech Cybersecurity, Version 2
- NEMA: ANSI/NEMA HN 1-2019 Manufacturer Disclosure Statement for Medical Device Security
- FDA: Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions
- US Code (LII): 21 U.S.C. 360n-2, Ensuring cybersecurity of devices
- HSCC: Medical Device and Health IT Joint Security Plan v2
- CISA: Known Exploited Vulnerabilities Catalog
Related NHI Mgmt Group resources: MDS2 and Medical Device SBOM Guide for Hospitals · Medical Device Security Guide · Hospital Cyber Security Programme Guide · Hospital Clinical Network Segmentation Guide. This template summarises model contract language and regulatory guidance and is not legal advice.